Perbandingan Stabilitas Bahan dan Kualitas Citra Grayscale pada Phantom Ultrasonography (USG) Berbasis Agar dan Oli
DOI:
https://doi.org/10.24252/jft.v12i2.57188Keywords:
Agar, Grayscale, oil, phantom, Ultranography (USG)Abstract
This study aims to compare the stability of materials and grayscale image quality of agar and oil-based ultrasonography (USG) phantoms. The method used includes making two types of phantoms with the same object, then imaging using a MINDRAY brand Ultrasography (USG) device type DP-10. Grayscale data was analyzed by calculating the average, standard deviation, and visual evaluation based on homogeneity, contrast, noise, and edge sharpness. The results showed that both phantoms had grayscale values that were still within the tolerance range of ±10% of the average, indicating the stability of wave propagation at an acceptable depth. However, the oil phantom showed a higher average intensity value of 9.838 cm and a lower standard deviation of 0.152053, compared to agar 9.584 cm and a standard deviation of 0.202. The oil phantom image appeared more homogeneous, sharp, and had minimal noise compared to the agar phantom. Thus, the oil-based phantom is considered superior in producing stable and representative grayscale image quality, and is more durable than agar
Downloads
References
[1] C. F. R. Putri, G. Maslebu, and A. Setiawan, “Analisis Densitas Larutan Berbasis Citra Phantom Agar dalam Kendali Mutu USG,” Positron, vol. 11, no. 1, p. 26, 2021, doi: 10.26418/positron.v11i1.44404.
[2] D. Pangestuti, G. Maslebu, and S. Trihandaru, “Perancangan Pengukuran Akurasi Jarak Menggunakan Phantom Silicone Pada Kontrol Kualitas Ultrasonografi dengan Teknik Komputasi,” J. Sains dan Edukasi Sains, vol. 2, no. 1, pp. 36–41, 2019, doi: 10.24246/juses.v2i1p36-41.
[3] N. F. Suryaningsih, B. S. Nirwana, P. Wahyu Wigati, and H. Saidah, “Hubungan Tingkat Pengetahuan Ibu Hamil Tentang Ultrasonografi Dalam Pemeriksaan Kehamilan Dengan Perilaku Pemanfaatan Ultrasonografi,” J. Mhs. Kesehat., vol. 4, no. 1, pp. 1–10, 2023, doi: 10.30737/jumakes.v4i1.3505.
[4] Yenni and A. T. Angka, “Hubungan Tingkat Pengetahuan Ibu Hamil Tentang Pemanfaatan Usg Di Klinik Wijaya Kusuma Makassar,” Public Heal. Med. J., vol. 2(1), 22–3, pp. 22–31, 2024.
[5] A. Suzuki et al., “Oil Gel-Based Phantom for Evaluating Quantitative Accuracy of Speed of Sound Measured in Ultrasound Computed Tomography,” Ultrasound Med. Biol., vol. 45, no. 9, pp. 2554–2567, 2019, doi: 10.1016/j.ultrasmedbio.2019.05.011.
[6] Y. Wang et al., “Fabrication of SEBS Block Copolymer-Based Ultrasound Phantom Containing Mimic Tumors for Ultrasound-Guided Needle Biopsy Training,” Ultrasound Med. Biol., vol. 48, no. 6, pp. 1143–1150, 2022, doi: 10.1016/j.ultrasmedbio.2022.02.020.
[7] T. Z. Pavan and M. Evertsson, “Oil-Based Gel Phantoms and the Speed of Sound,” Ultrasound Med. Biol., vol. 46, no. 11, pp. 3170–3171, 2020, doi: 10.1016/j.ultrasmedbio.2020.07.011.
[8] A. Scorza, F. Orsini, and S. Andrea Sciuto, “Use of phantoms and test objects for local dynamic range evaluation in medical ultrasounds: A preliminary study,” I2MTC 2017 - 2017 IEEE Int. Instrum. Meas. Technol. Conf. Proc., 2017, doi: 10.1109/I2MTC.2017.7969844.
[9] F. Noo, K. Hahn, H. C. Davidson, U. Rassner, and M. E. Heilbrun, “Influence of the grayscale on phantom-based image quality assessment in x-ray computed tomography,” Med. Imaging 2015 Image Perception, Obs. Performance, Technol. Assess., vol. 9416, p. 94160G, 2015, doi: 10.1117/12.2082914.
[10] C. Noor Santi, “Mengubah Citra Berwarna Menjadi Gray Scale dan Citra biner Rina,” Teknol. Inf. Din., vol. 16, no. 1, pp. 14–19, 2011.
[11] A. Prasetio, “Citra Digital Dan Algoritma Penerapannya,” pp. 4–16, 2021, [Online]. Available: http://dx.doi.org/10.31237/osf.io/7amr8
[12] M. S. Ummah, “Perbaikan Kualitas Citra Grayscale Dengan Metode Median Filter,” Sustain., vol. 11, no. 1, pp. 1–14, 2019, [Online]. Available: http://scioteca.caf.com/bitstream/handle/123456789/1091/RED2017-Eng-8ene.pdf?sequence=12&isAllowed=y%0Ahttp://dx.doi.org/10.1016/j.regsciurbeco.2008.06.005%0Ahttps://www.researchgate.net/publication/305320484_SISTEM_PEMBETUNGAN_TERPUSAT_STRATEGI_MELESTARI
[13] L. C. Cabrelli, F. W. Grillo, A. A. O. Carneiro, and T. Z. Pavan, “Copolymer-in-oil tissue-mimicking material with tunable acoustic properties,” IEEE Int. Ultrason. Symp. IUS, vol. 2016-Novem, pp. 8–11, 2016, doi: 10.1109/ULTSYM.2016.7728859.
[14] P. Chen et al., “Acoustic characterization of tissue-mimicking materials for ultrasound perfusion imaging research,” Ultrasound Med. Biol., vol. 48, no. 1, pp. 124–142, 2022, doi: 10.1016/j.ultrasmedbio.2021.09.004.
[15] A. Pratikno, A. Prastiwi, and S. Ramahwati, “Ragam Simpangan Baku serta Cara Menghitung Ragam Simpangan Baku,” pp. 5–7, Mar. 2020, doi: 10.31219/osf.io/q9umh.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2025 Kusma Dwi Yuristika, Imam Tri Harsoyo, Pramesti Kusumaningtyas

This work is licensed under a Creative Commons Attribution 4.0 International License.